A 4-20mA module can be added to any Blackcomb 5.1/6.1 UV system. This module allows the user to monitor the performance of their system. With a sensor installed, the module will output data showing the UV output over time. If there is no sensor installed, the output will always display 20mA, unless there is a major alarm on the controller (such as lamp failure). Follow the instructions below to add a 4-20 mA module to a UV system.
- With the system powered down, plug the male plug into the IEP on either the controller, sensor or another module containing an IEP.
- Connect the 4-20 mA module to the monitoring device that will be used using 16-22AWG wire. A power supply of 5-30 V/1A MAX is required.
- Power on system.
Troubleshooting
If the troubleshooting steps below do not resolve the issue, please submit a ticket for support. New Ticket
Issue: Connection Failure
Issue: Module Failure Alarm
- Unplug controller from outlet for 1 minute and plug it back in.
- If the issue has not been resolved, submit a ticket for support.
Issue: 4-20mA module not detected
If a 4-20mA module is installed and the screen below is displayed, the module is not being detected. Follow the steps below to resolve the issue.
Note: this screen is normal when there is no 4-20mA module installed.
- Unplug controller from outlet.
- Plug in module directly to controller and unplug any other modules/sensor.
- Plug controller into the outlet.
- Let the system power up and check for the screen below.
Issue: No output to remote monitoring device.
- Hold down button to ensure remote alarm module is detected.
- Ensure power is being supplied to the module. A power supply of 5-30 V/1A MAX is required.
- Check connections; tighten.
- Unplug system, unplug module and plug it in. Test separately.