You are using an unsupported browser. Please update your browser to the latest version on or before July 31, 2020.
close
You are viewing the article in preview mode. It is not live at the moment.
Home > Blackcomb > Troubleshooting > Blackcomb - Low UV Intensity Alarm
Blackcomb - Low UV Intensity Alarm
print icon

A low UV output alarm can only be displayed if a UV sensor is installed in a UV system. Many different factors can cause the UV intensity reading to be low when a UV sensor is installed. If low UV intensity is detected, one of the following screens will be displayed.

 

                    

 

When determining the cause of a UV alarm, the first step is to perform a water quality test. It is important that water quality parameters fall within the specifications below.

 

Hardness:                        <7 gpg (120 mg/L)

Iron (Fe):                          <0.3 ppm (0.3 mg/L)

Manganese (Mn):          <0.05 ppm (0.05 mg/L)

Turbidity:                         < 1 NTU

Tannins (organics):        <0.1 ppm (0.1 mg/L)

UVT (transmittance):    >85%

 

If the water quality falls outside the requirements, pre-filtration must be installed to resolve the issue. If the water quality falls within the requirements above, a dry test should be performed following the directions below.

Dry Test Procedure

  • Shut off the water source.
  • Unplug the controller from the power source.
  • De-pressurize the chamber and drain water.
  • Clean and dry the sensor and reinstall.  How to Clean a UV Sensor
  • Clean and dry the quartz sleeve and reinstall. How to Clean a Quartz Sleeve
  • With the water still shut-off, plug the controller into the power source.
  • Check the UV intensity. If 100%, the UV sensor is operating normally.

If the dry test shows a good result, other possible causes of low UV alarms are:

  • Air bubbles in water.
    • This can be tested by taking a water sample after the UV system in a clear glass. Look to see if bubbles form. If they do, there are air bubbles in the water. This must be purged from the water supply before the UV system.
  • High water temperatures.
    • This is seen when water is not flowing through the chamber for long periods of time. A flowmeter or a temperature relief valve can be installed to mitigate this issue.  

 

 

If the UV sensor is displaying a low UV intensity during a dry test, or if the cause for the alarm is undetermined after troubleshooting, please submit a ticket for support. New Ticket

 

Feedback
0 out of 0 found this helpful

scroll to top icon